10 – 23
Transpector SPS Operating Manual
10.6.2.5.2 Time Between Scans
Time Between Scans
can be set to a value greater than the minimum scan time
if additional wait time is desired. This field can only be set if
Manual
is chosen as
the
Delay Mode
.
Figure 10-40 Time between scans
NOTE: Time Between Scans
cannot be set lower than the minimum scan time.
10.6.2.5.3 Stabilization Time
Stabilization Time
is the amount of time FabGuard Explorer waits after the
Start
Mode
condition has been met before setting the Emission and EM to their specified
Start states and beginning data acquisition. For example, if the recipe starts based
on an analog input and an additional wait time is desired before the Emission is
turned on.
10.6.2.6 Correction
Defines the type of
Baseline
correction used by FabGuard Explorer (if any
correction at all) and whether or not
Peak Lock
is turned on.
10.6.2.6.1 Baseline
Sets the
Baseline
to
Off
or
Mono
.
For more information, refer to the FabGuard
Explorer Operating Manual.
Figure 10-41 Baseline