7 - 5
Transpector MPH Operating Manual
Poor
sensitivity
Sensor contaminated
Degas sensor.
Bake-out sensor.
Service sensor.
System pressure too low
Increase sample pressure, if
possible.
Mass calibration (resolution)
Adjust Mass Calibration,
increase peak width.
Sensor operating parameters
set wrong
Check settings of: electron
energy, ion energy, focus
emission current.
Improper calibration
Ensure that the total pressure
gauge used for sensitivity
calibration was properly
calibrated.
Electronics failure
Return to INFICON for repair.
EM has low gain
Bake-out sensor.
Replace sensor.
Poor peak shape
Sensor contaminated
Degas sensor.
Bake-out sensor.
Service sensor.
System pressure too high
Verify pressure less than
5E-4 Torr for
Transpector MPH.
Mass calibration required
Perform mass calibration.
Electronics failure
Return to INFICON for repair.
Table 7-1 Symptom—Cause—Remedy chart (continued)
SYMPTOM
CAUSE
REMEDY