Maintenance
The steel capillary may be removed in that way for the following sniffer tips:
3
Take out the steel capillary and remove the taper gasket at the top of the sniffer
tip.
Notice
The steel capillary can be cleaned with compressed air or a thin steel wire.
4
Reinsert the steel capillary and reinstall the sniffer tip.
5
Screw the plastic capillary filter onto the sniffer tip.
6
Please recalibrate the Protec P3000 for most accurate display of leak rate.
7.6.2
Switching from plastic to metal capillary filter
Notice
When you switch from the plastic capillary filter to the metal capillary filter,
do not forget to re-install the taper gasket as otherwise the SL3000 sniffer
line will be leaky.
1
Unscrew the plastic capillary filter.
2
Insert taper gasket (Fig.
/1)
3
Re-install the capillary in the sniffer tip.
4
Screw in metall capillary filter at the end of the sniffer tip.
5
Please recalibrate the Protec P3000 for most accurate display of leak rate.
Fig. 84
Steel capillary sticking out of sniffer tip flange
cat. no.
length
122 09
FT600
600 mm
flexible
122 13
ST312
120 mm
rigid
122 14
FT312
120 mm
flexible
122 15
ST385
385 mm
rigid
122 16
FT385
385 mm
flexible
122 18
FT200
200 mm
rigid
122 66
FT250
250 mm
flexible
122 72
ST500
500 mm
45° angled