Fig. 30: "Measure" window (presentation of results after calibration)
6.4.2 Connect external calibration leak
You have the option of connecting an external calibration leak to the device.
You cannot calibrate the device with the external calibration leak.
You can only calibrate the device with the internal calibration leak.
1
Place the external calibration leak with the copper sealing ring on the flange.
2
Tighten the 6 screws on the flange only slightly.
3
Recommendation: Using the modified Allen key 9/64 inch (included in delivery),
tighten the 6 screws on the flange according to the 2-2-3 system. For example:
Tighten screw 1. Tighten the screws in the order 3, 5, 2. Tighten the screws in
the order 4, 6, 3. Tighten the screws in the order of 5, 1, 4, etc. Repeat this
procedure until the gap between both flanges is almost completely closed.