INFICON
Installation | 5
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
35 / 100
1
To establish the compressed air supply for the device, insert the compressed air
hose into the air pressure quick coupling of the device until the quick-action
quick-release coupling engages.
2
Connect the other end of the compressed air hose to the compressed air
supply.
3
Set an operating pressure (relative pressure) > 7.5 bar and < 9 bar. An
overpressure valve is integrated, which, when the pressure applied to the device
is exceeded, audibly releases pressure.
4
Open the valve on the compressed air supply.
5.5 Connecting the purge gas
Use only Argon 5.0 or better as purge gas.
Use a metal line made of stainless steel or copper with a diameter of 6.35 mm for the
purge gas supply.
The sealing elements ferrel and clamping ring remain on the metal line and do not
have to be renewed as long as the same metal line is used.
► If you are using a new metal line for the purge gas supply, replace the ferrel and the
clamping ring of the compression fitting.
Compression fitting for the purge gas connection
Nut
Fig. 12: Connecting the purge gas
1
To install the purge gas supply for the device, plug the metal line of the purge
gas supply into the compression fitting of the device.
2
Using an open-ended wrench, hold the nut and tighten the union nut of the
compression fitting with an open-ended wrench and sufficient torque.
Summary of Contents for Pernicka 700H CHLD
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 99: ......