6 | Operation
INFICON
54 / 90
HLD6000-Operating-instructions-kinb43en1-05-(2107)
ð
If the device is switched off or the USB stick is removed from the USB
memory location without finishing the data record then the data for the current
hour being recorded is lost.
Stopping recording
1
> Recorder > Recorder settings
ð
Alternatively, when on the touchscreen select the button
.
2
Press the "Off" button under "Data Record".
3
Stop data recording by selecting the button
.
6.5.2 Evaluating measured data
A file with measurement data is structured as follows:
For example
// Record file: \L0000001.txt
// Created by HLD6000CU V0.11.02.18681
// HLD6000CU Ser.-No.: 00000000000
// HLD6000 Ser.-No.: 00000000000
// HLD6000MB V0.22.06(1.04.00)
// Probe V1.00
// Probe Ser.-No.: HLD5000 probe
// Probe Type: SMART (R134A)
// IO1000 V0.05.00(0.02.02)
// IO1000 Ser.-No.: 00000000000
// BM1000 not connected
// Start time: 23.06.2014 08:58:25
Time Leakrate[g/a] Status
0.0 1.82E-02 MEASURE
0.5 1.82E-02 MEASURE
1.0 4.16E-03 MEASURE
1.5 1.29E-02 MEASURE
2.0 0.00E+00 MEASURE
2.5 1.02E-02 MEASURE
3.0 1.75E-03 MEASURE
3.5 3.43E-03 MEASURE
4.0 3.43E-03 MEASURE
4.5 0.00E+00 MEASURE
5.0 8.20E-03 MEASURE
5.5 1.71E-02 MEASURE
6.0 1.52E-02 MEASURE
In the example shown the data record started on 06/23/2014 at 8:58:25. From this
start time onwards a measurement result was recorded every 500 milliseconds.
1.82E-02 means 1.82 x 10
-2
as the leakage rate per year.
The following status entries are possible:
Status entry
Meaning
UNKNOWN
Unknown, as, e.g., there is no
communication
RUNUP
Run up
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......