10.2.2 Event Log Files
Event Log Files are created automatically by the system to record system state
information, such as heating or valve changes, for future reference. These files are
created daily when running the HAPSITE using the PC software. These files are
saved in the Sensor directory Events folder (as shown in
) as
C:\haps4\
. The events would be saved in
C:\haps4\Events
by specifying
this location. The location where log files are saved can be changed only by
changing the Sensor directory.
By default, log files are deleted after 30 days. To modify how long the files are
saved, open up sensor properties by right clicking on a
Sensor
icon and selecting
Properties
; or by double clicking on a sensor; or by selecting the menu item
Tools
>> Sensor Properties
. Select the
Data Settings
tab. The window shown in
below will appear.
Figure 10-14 Sensor Properties - Data Settings
On the
Maintain Logs
line of Data Settings, the number of days which the log file
will be saved may be changed or to save all log files created indefinitely. If a
number of days is selected, each time HAPSITE Smart IQ is run the HAPSITE will
erase files which are older than the number of days selected.
NOTE:
This setting is saved in the HAPSITE Smart IQ.ini file in the HAPSITE
Smart IQ program install directory. If for any reason the HAPSITE.ini file is
reloaded or deleted, the setting will change back to the default setting of
Maintain Logs
for 30 days. Log files use the naming convention:
LogYYYYMMDD.evt, where YYYY is year, MM is month, and DD is day.
Summary of Contents for HAPSITE Smart
Page 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Page 2: ......
Page 12: ......
Page 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Page 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Page 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 548: ......