INFICON
Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
59 / 152
ð
You can listen to the set volume using the "Test" button.
3
If necessary, change the minimum volume.
ð
The "Minimum volume" is the volume for the audible alarm which cannot be
undershot. If you select a value greater than 0, the lowest volume level is
shown on the measurement screen after pressing the volume icon. This
allows you to prevent a non-authorized person from setting a value that is
below the ambient noise level.
ð
Setting range: 0 … 15
4
If you want to define the cause of an alarm or the type of notification beep more
accurately, then make your selection with the “Audio alarm mode” between
“Proportional leak rate”, “PINPOINT”, “SETPOINT”, “TRIGGER” and “Off”.
ð
"Proportional leak rate": The frequency of the audible signal is proportional to
the bar graph display or diagram height. The frequency range is 300Hz to
3300Hz.
ð
"PINPOINT": The sound of the acoustic signal changes its frequency within a
specific range of leak rates. This range extends from a decade below the
selected setpoint value to a decade above. Below this leak rate range, the
sound is constantly low, above the sound is constantly high.
ð
"SETPOINT": The pitch is proportional to the leak rate. The signal sounds if
the leak rate exceeds the set setpoint 1. See also "Setting setpoints [
ð
"TRIGGER": If the set setpoint 1 is exceeded, then a two-pitch signal is
issued. See also "Setting setpoints [
65]".
5
If you want to suppress the alarm for some time after pressing the START
button, then set “Audio alarm delay“. For example: Time period during the
pump down of a test chamber system.
ð
After pressing the START button the audible signal is activated as soon as
the leak rate falls below the setpoint value 1 or the alarm delay time expires.
This setting is only for the audible alarm type “SETPOINT“ and “TRIGGER“.
6
If you wish to have notification beeps, activate the option "Notification beep".
7
Save
.
6.2.9 Change the protection settings
You can protect the device from Helium contamination with gross leaks and from the
penetration of particles. Furthermore, you can change the maximum evacuation time.
ü
Supervisor
rights
1
> Setup > Measurement > Protection
2
If you want to switch-on the gross leak protection, activate the option "Gross
leak protection" and set under "Gross leak protection limit" a shut-down
threshold.