INFICON
Description | 4
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
31 / 152
4.3.4 Connections for accessories and control signals
SNIFFER
Power switch
Connection for
mains cable
Venting connection
Purge gas connection
ACCESSORIES
LD
REMOTE CONTROL
SERVICE
Fig. 6:
Connections for accessories and control signals
SNIFFER
For connecting the sniffer line SL3000
ACCESSORIES
Electrical connection of the sniffer line SL200
LD
Connection for bus module BM1000 or I/O module IO1000, maximum*
)
cable
length of INFICON data cable 30 m
REMOTE CONTROL
For connecting the remote control RC1000 via cable, maximum*
)
length 28 m,
or for direct connection of the radio transmitter via an adapter. The remote
control is not part of the scope of delivery of the device.
SERVICE
Connection only to be used by the Service
Connection for mains
cable
For power supply of the device
Venting connection
For hoses with an outside diameter of 14 mm, see also “Venting connection
Purge gas connection
For hoses with an outside diameter of 8 mm, see also “Purge gas connection
*) To avoid the display of incorrect measured values, the maximum cable lengths
specified must be observed.
NOTICE
The electronics of the device can be destroyed.
► Only connect devices that are provided for the appropriate interfaces.
The connections for the external devices show a safe disconnection from the power
supply and are within the range of the safety extra low voltage (SELV).
Summary of Contents for Fab UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Page 151: ......
Page 152: ......