6 Operation
WARNING
Risk of injury from flammable solvents
The solvent in the calibration leak is highly flammable.
Overheating can damage the membrane, solvent can leak and ignite from an ignition
source.
► Observe the manufacturer's safety data sheets and follow the applicable work
instructions.
► Avoid heating the calibration leak to high temperatures.
NOTICE
Health risk due to gases and vapors
Operation of the leak detector may produce hazardous vapors.
► Connect the Gas Detection Unit and the Gas Control Unit to an exhaust line.
► Avoid inhaling harmful gases or vapors.
► Ensure sufficient ventilation at the installation location.
► Observe the safety instructions in the safety data sheets for the test objects.
► Provide an installation location where blockage of the exhaust lines is not possible
or can be detected.
► Provide an installation location with adequate ventilation or alternatively an
installation location where air quality is tested and monitored for harmful
substances.
NOTICE
Property damage due to overheated device
The Gas Detection Unit as well as the Gas Control Unit will become warm during
operation and may overheat without adequate ventilation.
► Keep the underside of the gas control unit unobstructed.
► Do not block the ventilation opening for the filter.
► Make sure that there is adequate ventilation at the Gas Detection Unit: Free space
at least 20 cm on the sides, at least 10 cm at the front and rear.
► Keep heat sources away from the battery leak detector.
► Do not expose the battery leak detector to direct sunlight.
► Please note the technical specifications.