Equipment Settings
4-7
ki
n
a
22
e c
h
ap
te
r 4
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
4.4.3
Selecting trigger values and units of measurement
In the “
EDIT
GAS
” info page the
TRIGGER
/
UNIT
sub-menu page will open by selecting
the “
TRIGGER
” line item. You can change the trigger level setting by the use of the left
“up” and “down” keys. The correct settings must be confirmed by the “
OK
” button.
You can leave the sub-menu without making any changes by using the “
ESC
” key.
The ranges the trigger can be set to is listed in the following table.
Default value: 4.0
The leak rate unit can be changed by the use of the right “
UP
” and “
DOWN
” keys.
Available leak rate units include mbar l /s, Pa m³ / s, Torr l / s, atm cc / s, g/a, oz/yr,
ppm.
Default unit: g/a
Unit
Lower trigger limit
Upper trigger limit
g/a
0.1
1,000.0
oz/yr
0.004
100.00
ppm
1
999999
mbar l/s
2x10
-7
9.9x10
-2
Pa m³/s
2x10
-8
9.9x10
-3
atm cc/s
2x10
-7
9.9x10
-2
Torr l/s
2x10
-7
9.9x10
-2
Fig. 4-8 Setting-up trigger level + unit of measurement
Pos. Description
Pos.
Description
1
Increasing the trigger value
3
Changing the unit of measurement
2
Decreasing the trigger value
4
Changing the unit of measurement
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...