For Ecotec E3000RC only
The Ecotec E3000RC has no built-in display unit but a connectors plate is mounted
instead. Please connect the external display unit with the 5 m connecting table (Cat.-
no. 551-102).
1.6.3
Electrical Connections
Notice:
The local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Ecotec E3000
can be read off from the name plate left of the power switch. The mains
voltage setting of the Ecotec E3000 is fixed and can not be changed. A
separate fuse for each of the mains conductors has been integrated into
the mains socket (
Fig. 2-1
/6).
Fig. 1-8 Ecotec E3000RC with external display unit for: (a) bench top use (left side),
(b) rack mounting (right side)
Fig. 1-9 Electrical connections
Pos. Description
Pos. Description
1
Headphone port
4
Name plate
2
I/O Port
5
Power switch
3
RS232 interface
6
Power connector
Warning
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Ecotec E3000 with the ground conductor unconnected is not
permissible.
1
2
3
4
5
6
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...