3 | Specifications
INFICON
6 / 357
074-710-P16A D-TEK 3 Operating Manual
3 Specifications
Usage
indoor/outdoor
Sensor type
infrared
Compatible refrigerants
• Refrigerant sensor
(standard, PN 724-701-G1)
All CFCs, HCFCs, HFCs, HFOs, and
blends (including A2Ls)
• CO
2
sensor (PN 724-701-G2)
R744 (carbon dioxide)
• Flammable refrigerant (HC) sensor
(PN 724-701-G3)
R290, R600a, R441a (see safety
warnings)
Minimum sensitivity (Pinpoint mode,
Super sensitivity)
0.03 oz/yr (1 g/yr)
1
Battery type
lithium ion
Charging input type
micro USB
Charging time (starting at 0%)
approximately 3 hours
Battery life
approximately 9 hours
Input voltage
5 V (dc) ±5%
Input current
1 A ±5%
Warm-up period
45–90 seconds
Temperature ranges and humidity
• Storage
-20–60°C (-4–140°F)
• Operating
2
-20–50°C (-4–122°F)
• Charging
0–45°C (32–113°F)
• Humidity
95% RH NC maximum
Altitude
2000 m (6500 ft.)
Pollution degree
2
Overvoltage category
2
Weight (with battery; not including
carrying case or accessories)
0.97 lb. (0.44 kg)
1
To achieve optimal performance and the specified sensitivities, it is recommended to
allow D-TEK 3 to run for 15 minutes prior to use.
2
Use in temperatures below 0°C (32°F) should be limited. Extended warm-up time is
recommended before use in low temperature environments.
English
English
Summary of Contents for D-TEK 3
Page 135: ...INFICON 符合性声明 1 074 710 P16A D TEK 3 Operating Manual 135 357 1 符合性声明 中文 中文 ...
Page 157: ...INFICON 法令遵守に関する宣言 1 074 710 P16A D TEK 3 Operating Manual 157 357 1 法令遵守に関する宣言 日本語 日本語 ...
Page 308: ...1 적합성 선언 INFICON 308 357 074 710 P16A D TEK 3 Operating Manual 1 적합성 선언 한국어 한국어 ...
Page 356: ......