Operation | 6
Contura® S400-Operating-instructions-jina91en1-09-(2001)
39 / 68
2
> Data backup > Save data
6.8 Restore user and product data
Overwrite current user and product data
By restoring stored user and product data the current state of these data is deleted.
ü
Supervisor - rights
ü
You have a USB flash drive (FAT32 formatted) with user and product data from a
backup.
1
Insert the USB stick into a USB port on the device.
2
> Data backup > Restore data
6.9 Restoring factory defaults
You can restore the device to factory settings.
Loss of settings and measurement data
After resetting to factory defaults only the manufacturer factory settings is in the
memory of the device.
Backup important settings and measurement data beforehand on a USB flash drive.
See "Save user and product data [
38]" and "Transferring measurement data [
35]".
ü
Supervisor - rights
►
> Reset device
6.10 Calibrating the device
6.10.1 Reasons for calibration
In the following cases, calibration is mandatory:
• The installation site of the leak detector is located above 1000 m above sea level.
• The measuring chamber was changed.
• There are operational requirements for a calibration.
6.10.2 Calibration means
There are various calibration means for the device available: