> General > Display (Tab) > Auto-rotate screen
ð
The symbols
(top right) and
(bottom right) appear.
2
Select auto-rotate screen press button next to crossed box
.
ð
To switch off the auto-rotate screen, press button next to empty box
.
6.11 Switching OFF
Switching OFF
1
Press the "On / Off” key for 3 seconds.
ð
The leak detector asks whether you want to switch off the instrument.
2
Select “Yes” and press the check mark key.
Hard switch off
Some errors can make the leak detector go into watchdog state. The acoustic alarm
beeps continuously and the status LED is red. The leak detector may not even be
reacting to the menu buttons. If this happens you must turn the leak detector off. Press
and hold the "On / Off” key until the detector shuts down (roughly 9 s).
Do not use this option without cause. You risk losing settings and the bias of the
oxygen sensor on SXT and SXGT models. This leads to a prolonged warm up at next
power on. It can take up to 12 hours to stabilize the oxygen sensor if sensor bias is
lost form more than a few minutes. If you need to do a hardware switch off you should
restart the leak detector immediately, If the leak detector malfunctions again you will
need to contact INFICON service.
1
Press the "On / Off” key for 9 seconds.
2
Restart the leak detector as soon as possible to restore O
2
sensor bias (IRwin
SXT and SXGT).
6.12 Warnings and error messages
Warnings
Warnings are given for conditions that can impair the accuracy of measurements. The
operation of the leak detector is normally not interrupted but some functions may be
barred or indicated as non-compliant by status LED being yellow and values being red
instead of black.
If the navigation button “Menu“
is colored red this means that there has been an
error that you have not yet checked. If the cause of the error remains there will be a
new warning/error alarm and a new entry in the error log. The navigation button turns
blue again when you have read the warnings list.
Open the warnings log list:
►
>
> Warnings
Error messages
Errors are events that force the interruption of the operation. Errors are also logged in
the warnings log list.
Summary of Contents for 580-000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 99: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 99 128 ...
Page 101: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 101 128 ...
Page 102: ...9 Certificates INFICON 102 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 103: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 103 128 ...
Page 105: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 105 128 ...
Page 110: ...9 Certificates INFICON 110 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 111: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 111 128 ...
Page 113: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 113 128 ...
Page 114: ...9 Certificates INFICON 114 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 115: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 115 128 ...
Page 116: ...9 Certificates INFICON 116 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 117: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 117 128 ...
Page 119: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 119 128 ...
Page 120: ...9 Certificates INFICON 120 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 121: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 121 128 ...
Page 122: ...9 Certificates INFICON 122 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 123: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 123 128 ...
Page 126: ...Index INFICON 126 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 127: ......
Page 128: ......