6.10.1 Update the user interface software
Import software updates using a USB flash drive.
NOTICE
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated.
ü
Supervisor rights
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to a USB port on the device.
3
> Update > Update operating unit
ð
At the top in the window the active software version of the user interface is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
ð
After completion there is an automatic restart of the operating unit.
6.10.2 Updating the software of the basic unit
Import software updates using a USB flash drive.
NOTICE
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated.
ü
Supervisor rights
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
> Update > Update basic unit
ð
At the top in the window the active software version of the basic unit is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......