4. Select the correct serial port. A serial-to-serial adapter is typically COM1. A USB-to-
serial adapter is typically a different COM number.
Input Configuration
1. Click General information, parameters, and options.
2. Click the Options tab.
l
For a digital or pulse input: enter a Parameter name, value of a parameter,
and parameter units under Input 1.
l
For an analog input: enter a Parameter name, high and low values, and
parameter units under Input 2.
The
Power supply to the probes
checkbox must be checked for
analog input devices.
For HydroVu users, the Parameter name and units are directly
passed to the cloud. They cannot be changed from HydroVu.
3. Click Write options and parameters.
4. Close the window. Click Probe and set-point configuration.
5. From an open probe drop down menu, select Digital in, Pulse in, or Analog in.
6. Click Write Probe and set-points.
7. Close the window.
If you require alarming for the input device, continue with steps 8-12.
8. Click theParameters monitoring tab.
9. Select the input from one of the probe drop down menus.
10. Select the parameter from the parameter drop down menus.
11. Set the Warning and Alarm levels for the parameter, and set the level mode.
12. Click Write Probe and set-points.
Disconnect the communication cable and remember to connect
the stripped-and-tinned cable of the device to the Tube before
conducting any testing or when deploying the device.