4.4.
D
EVICE
S
TATUS
In detection mode press the button once and
the screen will show the device status
including max value, min. value, STEL value,
TWA value, current time, serial number and
version number.
4.5.
C
ALIBRATION
1.
Gas input
2.
Calibration cover
3.
Fastener
With the device is in detection mode hold the
button for more than 20 seconds. The screen
will then turn off and then back on. When the
screen shows calibration mode release the
button.
The below screen will be displayed:
1.
Entering the calibration interface
2.
Sensors warming up
3.
Zero calibration
Ensure the instrument is in “fresh air”
4.
Gas input
The screen will display calibration gas
concentrations. After applying the standard
gas, the detector will start to analyze.
5.
Auto calibration
6.
Calibration pass/fail