Page 31
NOTE
: The included tips must be autoclaved prior to use as they do
not come pre-sterilized.
NOTE
: Ensure that the surface of the mirror does not show residues,
smudges, scratches, or any damage, as this would affect the perfor-
mance of the device.
6.
Maintenance
6.1
Cleaning the Handpiece
The entire body, cord and base of the scanner must be wiped down using a Federal
Environmental Protection Agency (EPA) approved disinfectant that is labeled and
specified for tuberculocidal/ mycobactericidal activity. Do not use disinfectant on the
nozzle.
Recommended and approved surface disinfectants:
•
Birex Wipes: TB Claim=10 minutes- ‘Phenolic (Dual) Water-Based’
•
Prospray Wipes: TB Claim=10 minutes- ‘Phenolic (Dual) Water-Based’
•
Cavicide Wipes: TB Claim=3 minutes- ‘Phenolics (Dual) Alcohol-Based’
•
DisCide ULTRA Wipes: TB Claim-1 minute- ‘Phenolics (Dual) Alcohol-Based’
•
Maxiwipe Germicidal Cloth: TB Claim=5 minutes- ‘Phenolics (Dual) Alcohol-Based’
•
Ster 1 Plus: TB Claim=5 minutes- ‘Quaternary ammonium and Alcohol-Based’
6.2
Cleaning and Sterilization of Tip
NOTE
: All components of the scanner (excluding the tips) must be
wiped down and not sprayed. Avoid getting any moisture, alcohol or
disinfectant inside the open scanner chamber.