5.3 Connecting a Foot Pedal Switch
The scan sequence can be started with optional available foot pedal switches.
At the back side are two jack plugs to which the foot pedal switches can be connected.
The jack plugs are labeled with “FS1” and “FS2”.
Notice:
Switch off the BOOKEYE Color
before
connecting or disconnecting the foot
pedal switch to the jack plugs.
5.4 Possible Disturbance at Power Up
After connecting the BOOKEYE Color to the mains voltage and turning the main power
switch on, the lamps should light up for a short moment.
This is a design-related behavior of the lamps electronic ballast.
Situation: When starting the device using the START button, one of the two lamps in the
lamp housing does not light up.
The lamp is not defective. A safety feature in the lamps electronic ballast prevents the
lamp from turning on. To resolve this situation, employ one of the two solutions, or both, if
necessary.
Important:
For both solutions the following is valid:
Do not
turn off the main power
switch and
do not
disconnect the device from the mains voltage.
Suggested solution 1:
After completing the device system start-up sequence, switch the device off using the
STOP button. Chapter 6.2 describes all necessary steps.
After a short waiting period of approx. 10 minutes, the device can be restarted via the
START button. Both lamps in the lamp housing should light up now.
Suggested solution 2:
After completing the device system start-up sequence, switch the device off using the
STOP button. Chapter 6.2 describes all necessary steps.
Unplug the lamp which did not light up from the lamp socket and plug it in again.
Note:
The lamps can be plugged into the lamp sockets or removed from the lamp
sockets without contact of live units.
Picture 4: Lamp housing. Pull in arrow direction to unplug the lamps.
Operation Manual
Page 17