Operation
6.4.4 Menu
item:
„Calibrate“
Calibration
When SEARCH
(lower point)
is activated, evacuation is to the lower point. If the pressure
does not fall any further, the value can be entered by using a comparison measurement de-
vice and confirmed with OK. The system is vented and the ambient barometric pressure en-
tered as the upper point. Do not confirm until the value has been established.
If the controller is integrated in a complete system or pump system (LVS), evacuation and
venting take place automatically when the stated buttons are pressed.
The calculated correction values are displayed and can be stored with Save.
CAUTION !
Ensure that the set pressure is also present at the sensor.
Designation
ILMVAC-Control
Control
panel
Explanation
lower point
CL
At low pressure - pressure of comparison measurement device
(at lowest possible pressure, e.g. 10 mbar)
upper point
CH
At normal pressure (high) - barometric pressure
113184
29