Process Status
The MiniSeq Control Software lists the status of at least 3 runs in the system Temp folder. From the Manage
Instrument screen, select Process status.
For each run name, the system lists the status of the following components:
u
Real-Time Analysis (RTA)—Based on the processing of BCL files
u
Local Run Manager—If Local Run Manager was used for the run
u
File Copy—Based on file transfer using the Run Copy Service
u
BaseSpace—If BaseSpace was used for the run
Sequencing Archive Folder
The MiniSeq Control Software saves run summary files on the system computer at D:\Illumina\MiniSeq
Sequencing Archive for each run performed on the instrument.
In this folder, a subfolder exists for each run performed on the instrument contains the following files:
u
RunCompletionStatus.xml—Contains the completion status, run folder name, number of cycles planned
and performed, cluster density, clusters passing filter, and estimated yield for the run.
u
RunParameters.xml—Contains information about run parameters and run components. Information
includes the RFID, serial number, lot number, and expiration date.
Automatic Check Errors
If errors occur during the automatic pre-run check, use the following recommended actions to resolve the
error.
If a pre-run check fails, the reagent cartridge RFID is not locked and can be used for a subsequent run.
However, the RFID is locked after the foil seals have been pierced.
System Checks
Recommended Action
Doors closed
Make sure that the compartment doors are closed.
Consumables loaded
Consumable sensors do not register. Make sure that each consumable is properly loaded.
On the run setup screens, select Back to return to the loading step, and repeat run setup.
Required software
Critical components of the software are missing.
Perform a manual software update to restore all software components.
Instrument disk space
The instrument hard drive does not have sufficient disk space to perform a run.
Clear run data from the instrument hard drive.
Network connection
The connection to the specified location of the output folder has been interrupted.
Although the check is labeled network connection, the system checks the connection to
any specified location of the output folder on a server, an external hard drive, or the local
drive.
Check the status of the connection to the specified output folder location.
Network disk space
The specified location of the output folder is full. Although the check is labeled network
disk space, the system checks any specified location of the output folder on a server, an
external hard drive, or the local hard drive.
Clear disk space from the specified output folder location.
Document # 1000000002695 v02 Material # 20014309
For Research Use Only. Not for use in diagnostic procedures.
39
MiniSeq System Guide