
A
One stripe is visible between the first clamp and the alignment bar.
B
Glass back plate pressed against alignment bar.
C
Stripes are not visible between the second clamp and the barcode.
9
Remove the assembled flow-through chamber from the alignment fixture.
10 Starting at the nonbarcode end, trim the spacers from each end of the assembly using scissors.
A
Trimming the spacer at the nonbarcode end.
B
Trimming the spacer at the barcode end.
11 Leave assembled flow-through chambers on the lab bench in a horizontal position until ready to load onto
chamber rack in the
Extend and Stain BeadChips
step.
u
Do not place on absorbent paper.
u
Do not place in the chamber rack until instructed to do so.
12 Wash the hybridization chamber reservoirs with DI H
2
O.
Immediate and thorough washing ensures complete removal of PB1 from the wells.
Extend and Stain BeadChips
This step washes unhybridized and nonspecifically hybridized DNA samples from the BeadChips, adds
labeled nucleotides to extend primers hybridized to the sample, and stains the primers. After the flow-
through chambers are disassembled, the BeadChips are coated for protection.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
58
Infinium HD Super Assay Reference Guide