C
lu
s
te
ri
n
g
W
it
h
o
u
t
S
a
m
p
le
T
ra
c
k
in
g
58
Material # 20004364
Document # 15065681 v01
Perform a Post-Run Wash
1
Wash the plate on the thermal stage with deionized water to remove any salts. Dry
with a lint-free cleaning tissue.
2
Fill the wash reservoir with approximately 12 ml deionized water and close the
instrument lid.
3
Select the checkbox to indicate that water is present, and then select
Wash
.
4
When the wash is complete, blot any excess water remaining in the wash reservoir.
Avoid the outlet ports to prevent fibers from entering the holes.
5
Select the checkbox to indicate that the wash reservoir is dry, and then select
Exit
.
The Start screen opens and your cBot 2 is ready for another run.
Summary of Contents for cBot
Page 4: ...iv Material 20004364 Document 15065681 v01 ...
Page 16: ...10 Material 20004364 Document 15065681 v01 ...
Page 22: ...16 Material 20004364 Document 15065681 v01 ...
Page 66: ...60 Material 20004364 Document 15065681 v01 ...
Page 76: ...70 Material 20004364 Document 15065681 v01 ...
Page 88: ...Index 82 Material 20004364 Document 15065681 v01 ...