iHealth BP7 Owner'S Manual Download Page 14

 

 

13

nearby electronic equipment. 

RF emissions 
CISPR 11 

Class B 

The monitor is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes. 

Harmonic emissions 
IEC 61000-3-2 

Class A 

Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3 

Complies 

Table 2 

For all ME EQUIPMENT and ME SYSTEMS 

Guidance and manufacturer’s declaration - electromagnetic immunity 

The Wireless Blood Pressure Wrist Monitor is intended for use in the electromagnetic environment 
specified below. The customer or the user of the Wireless Blood Pressure Wrist Monitor should assure 
that it is used in such an environment. 

IMMUNITY 

test 

IEC 60601test level 

Compliance level 

Electromagnetic environment - 

guidance 

Electrostatic 
discharge 
(ESD) 
IEC 61000-4-2 

± 6 kV contact 
± 8 kV air 

± 6 kV contact 
± 8 kV air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative 
humidity should be at least 30 %. 

Electrical fast 
transient/burst 
IEC 61000-4-4 

± 2 kV for power 
supply lines 
 

± 2 kV for power 
supply lines 
 

Mains power quality should be that of 
a typical commercial or hospital 
environment.   
 

Surge 
IEC 61000-4-5 

± 1 kV line(s) to 
line(s) 
± 2 kV line(s) to 
earth 

± 1 kV line(s) to 
line(s) 
± 2 kV line(s) to 
earth 

Mains power quality should be that of 
a typical commercial or hospital 
environment. 

Voltage dips, 
short 
interruptions 
and 
voltage 
variations 
on power 
supply 
input lines 
IEC 61000-4-11 

<5 % U

T

 

(>95 % dip in U

T

for 0.5 cycle 
40 % U

T

 

(60 % dip in U

T

for 5 cycles 
70 % U

T

 

(30 % dip in U

T

for 25 cycles 
<5 % U

T

 

(>95 % dip in U

T

for 5 s 

<5 % U

T

 

(>95 % dip in U

T

for 0.5 cycle 
40 % U

T

 

(60 % dip in U

T

for 5 cycles 
70 % U

T

 

(30 % dip in U

T

for 25 cycles 
<5 % U

T

 

(>95 % dip in U

T

for 5 s 

Mains power quality should be that of 
a typical commercial or hospital 
environment. If the user of the monitor 
requires continued operation during 
power mains interruptions, it is 
recommended that the monitor be 
powered from an uninterruptible 
power supply or a battery. 

Power 
frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical 
commercial or hospital environment. 

NOTE: U

T

 is the a.c. mains voltage prior to application of the test level. 

Table 3 

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Summary of Contents for BP7

Page 1: ...OWNER S MANUAL...

Page 2: ...ARTS AND DISPLAY INDICATORS 3 SET UP REQUIREMENTS 3 SET UP PROCEDURES 4 MEASUREMENT PROCEDURES 4 SPECIFICATIONS 6 GENERAL SAFETY AND PRECAUTIONS 6 BATTERY HANDLING AND USAGE 9 TROUBLESHOOTING 10 CARE...

Page 3: ...ystolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of the cuff circumference...

Page 4: ...ne and iPad models iPod touch 4th generation iPod touch 3rd generation iPhone 4S iPhone 4 iPhone 3GS iPad 3rd generation iPad 2 iPad The iOS version of these devices should be V4 0 or higher SET UP PR...

Page 5: ...DURES Apply the Cuff a Place your hand palm side up in front of you and wrap the cuff around a bare wrist 1 2 cm above the wrist joint b If the monitor is correctly placed the iHealth logo will be fac...

Page 6: ...lthcare professional for interpretation of blood pressure measurements Measuring without an iOS Device Offline measurement is enabled automatically at your first online measurement To start offline me...

Page 7: ...quency Band 2 402 2 480 GHz 13 Environmental temperature for operation 5 40 41 104 14 Environmental humidity for operation 90 RH 15 Environmental temperature for storage and transport 20 55 4 131 16 E...

Page 8: ...but the results may be inaccurate Please consult your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of puls...

Page 9: ...rsonnel could result in a hazard such as a fire or explosion Do not plug or unplug the power cord into the electrical outlet with wet hands If the AC adapter is abnormal please change the adapter Do n...

Page 10: ...avoid speaking or movement during the test Bluetooth connection unstable Bluetooth connection unsuccessful BP7 is abnormal or strong electromagnetic interference is present Reset iOS device Reset BP7...

Page 11: ...will possibly damage your unit 12 It is recommended that if the cuff is used for example in a hospital or a clinic it be disinfected twice a week Wipe the inner side the side that contacts skin of th...

Page 12: ...Lotus Global Co Ltd 15 Alexandra Road London UK NW8 0DP Tel 0044 20 75868010 Fax 0044 20 79006187 ANDON HEALTH CO LTD No 3 Jinping Street Ya An Road Nankai District Tianjin 300190 China Tel 86 22 6052...

Page 13: ...le fonctionnement OTHER STANDARDS AND COMPLIANCES The Wireless Blood Pressure Wrist Monitor corresponds to the following standards IEC 60601 1 2006 Medical electrical equipment Part 1 General require...

Page 14: ...kV for power supply lines 2 kV for power supply lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth...

Page 15: ...watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a...

Page 16: ...and mobile RF communications equipment transmitters and the monitor as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter...

Reviews: