STERYLIS ULTRA Operating Manual
Parameter code
Description
t200
AUTO-TEST function
t201
OZONE-SENSOR-TEST function
The individual parameters are changed using the „+” and „-” buttons in the „O
3
sterilisation” section. The
code of the selected parameter is displayed in the „O
3
sterilisation” section. To move to a given function,
select it and press and hold the „PROG.SET” button. The selection of a given function will be confirmed by
a short beep and all displays will blink.
Other functionality of the device
AUTO-TEST function
OZONE-SENSOR-TEST function
Other remaining functionality of the device is described in subsections 6.6.1, 6.6.2 and 6.6.3. These are:
AUTO-TEST function and OZONE-SENSOR-TEST function
To access the above mentioned functions, press and hold the „SET” button in the „Ozone time”
section of the device standby mode. Then enter the appropriate code to move to the above mentioned
functions.
Entering the access code:
The 4-digit code entered is shown on the display under „O3 sterilisation”. The „+” and „-” buttons change
the value by 1 active digit. An active number is the one that flashes at a high frequency, inactive numbers
light up constantly. The „PROG.SET” button changes the active digit to the next one on the right. After
entering the correct code, confirm it by pressing the yellow button „START” longer. If the code is incorrectly
entered, the code „XXXX” or „HHHH” will be displayed and a short beep will occur. In this case, the code
must be re-entered.
The available functions are under the following codes:
The function consists in an automatic check of the operating status of all actuators (fan, UV-C lamps,
ozone generators) and the ozone sensor. If the device detects an error in one of the elements during the
process, it will display the error. If an error is detected during previous use of the unit, the AUTO-TEST
function can erase this error if the unit operates correctly during this function (error erasing does not apply
to the ozone sensor).
During the AUTO-TEST function, the message „tESt” is displayed and the operating mode, warning
and safe ozone concentration lights blink quickly.
To interrupt the AUTO-TEST function, press any of the „START” buttons corresponding to filtration,
disinfection or sterilisation modes.
When the AUTO-TEST function is completed, the unit will go into standby mode.
The function consists of an automatic check of the ozone sensor’s operating status. If the device detects
an error in the ozone sensor during the process, it will display the error. If an ozone sensor error has been
detected during previous use of the unit, the OZONE-SENSOR-TEST function can erase this error if the
unit will function correctly during this function.
During the AUTO-TEST function: the message „O3_t” on the display in the „O3 sterilisation” section,
the measured instantaneous voltage value on the display in the „O3 concentration” section, the time
remaining for checking the ozone sensor on the display in the „Ozone time” section and the warning lights
and the safe ozone concentration are blinking quickly.
To interrupt the AUTO-TEST function, press any of the „START” buttons corresponding to filtration,
disinfection or sterilisation modes.