20
FE = window; FH = upper value; FL = lower value�
For this unit, the assignment of the parameters [dSx] and [drx] to the set and reset
points is strictly designed according to the VDMA guideline�
9.5.2 Set damping for the switching signal
►
Select [dAP] and set the damping constant in seconds (
T
value 63 %);
setting range 0�000���4�000 s�
Damping [dAP] affects the switch point / process data flow (IO-Link
communication) and the display�
9.5.3 Set damping for the analogue output
►
Select [dAA] and set the damping constant (rise time 10���90 %) in
seconds; setting range 0�000���4�000 s�
Damping [dAA] only influences the analogue output / analogue
signal path�
9.5.4 Reset all parameters to factory setting
►
Select [rES]�
► Press [●].
► Press [▲] or [▼] and keep pressed until [----] is displayed.
► Briefly press [●].
We recommend noting down your own settings before carrying out a reset
Also the operating mode [CMPT] is reset to the factory setting
([CMPT]=[2])�
9.5.5 Set colour change of the display
►
Select [coLr] and set the function:
- [rEd] = display colour red (independent of the measured value)�
- [GrEn] = display colour green (independent of the measured
value)�
- [r1ou] = display colour red when OUT1 switches�
- [G1ou] = display colour green when OUT1 switches�
- [r-cF] = display colour is red if the measured value is between
the freely definable limits [cFH*
)
and [cFL]*
)
�
- [G-cF] = display colour is green if the measured value is between
the freely definable limits [cFL]*) and [cFH]*)�
*
)
The parameters [cFH] and [cFL] can only be selected in the menu tree if
[r-cF] or [G-cF] has been activated�