12.1 Operation indication
[----] (continuous)
Initialisation phase after power on
[numerical value] +
LED 1
Current level in cm
[numerical value] +
LED 2
Current level in inches
LED 7
Switching status OUT-OP:
• LED on: No overflow (overflow prevention point OP not
reached)
• LED off: Tank is full (overflow prevention point OP is reached)!
LED 8
Switching status OUT1 (on when switching output OUT1 is
switched)
[----]
Level below the active zone
[FULL] + [numerical
value] alternately
The overflow prevention OP is reached (overflow warning) or the
level is above the active zone
Adjustment [cOP] of the overflow prevention OP necessary
[Loc]
Unit locked via operating keys; parameter setting not possible
For unlocking press the two setting buttons for 10 s�
[uLoc]
Unit is unlocked / parameter setting is possible again
[C�Loc]
The unit is temporarily locked; parameter setting via IO-Link is
active (temporary locking)
[S�Loc]
Unit is permanently locked via software� This locking can only be
removed with a parameter setting software�
[CodE]
The unit is password protected� Before the unit can be
configured, the correct password must be entered� Use
parameter [KEy�C] to enter the password
12.2 Read the set parameters
►
Briefly press [Mode/Enter] (if required, repeat several times)�
> Menu structure is scrolled until the required parameter has been reached�
►
Briefly press [Set]�
> Respective parameter value is displayed for 30 s�