LI213x
Binary level sensor for overflow prevention / leakage detection according to WHG
20
7.8 Parameter setting examples via IO-Link
7.8.1 Application as overflow prevention or maximum monitoring
u
Set switching output OUT2 to temperature detection.
Example: [SEL2] = [TEMP]
u
The temperature output OUT2 shall be employed as alarm output (as normally closed with a high
temperature threshold). Example: [ou2] = [Hnc]
u
Set the alarm temperature to 80°C.
Example:
[SP2 (FH2)-TEMP] = 80;
[rP2 (FL2)-TEMP] = 75.
u
Transfer the parameter data to the device
u
Carry out the empty adjustment by executing the system command [Empty Teach].
u
If possible, carry out a full adjustment to the medium to be detected:
Fill the tank until the active zone of the device is completely covered.
Notes:
Ò
Empty adjustment and full adjustment.
u
Execute the system command [Full Teach].
Any other settings are left at factory setting.
7.8.2 Application as leakage detection
u
Make sure that the collection system is empty, otherwise empty it completely.
u
Carry out the empty adjustment by executing the system command [Empty Teach].
u
If possible, carry out a full adjustment to the medium to be detected:
Fill the collection system until the active zone of the device is completely covered.
Notes:
Ò
Empty adjustment and full adjustment.
u
Execute the system command [Full Teach].
Any other settings are left at factory setting.