21
UK
9.3 Settings for consumed quantity monitoring
9.3.1 Configure quantity monitoring via pulse output
►
Select [OU1] and set [ImP]�
►
Select [ImPS] and set the volumetric flow quantity at which 1 pulse is
provided (→ 9.7).
►
Select [ImPR] and set [YES]
> Pulse repetition is active� Output 1 provides a counting pulse each time
the value set in [ImPS] is reached�
9.3.2 Configure quantity monitoring via the preset counter
►
Select [OU1] and set [ImP]�
►
Select [ImPS] and set the volumetric flow quantity at which output 1
switches (→ 9.7).
►
Select [ImPR] and set [no]
> Pulse repetition is not active� The output switches ON if the value set in
[ImPS] is reached� It remains switched until the counter is reset�
9.3.3 Configure program-controlled meter reset
►
Select [rTO] and continue with a) or b)�
a) Reset the meter manually:
►
Press [SET] until [rES�T] is displayed, then briefly press [Mode/Enter]�
b) Enter the value for time-controlled reset
►
Press [Set] until the requested value is displayed (intervals from 1 hour
to 8 weeks), then briefly press [Mode/Enter]�
9.3.4 Deactivate meter reset
►
Select [rTo] and set [OFF]
The meter is only reset after overflow (= factory setting)�
Overflow: After the maximum value (9 999 999 Nm³) the meter is reset to 0�
9.3.5 Configure meter reset using an external signal
►
Select [OU2] and set [InD]
►
Select [Din2] and set the reset signal
-[HIGH] = reset for high signal
-[LOW] = reset for low signal
-[+EDG] = reset for rising edge
-[-EDG] = reset for falling edge