10.4.5 Time-controlled counter-reset
► Select [rTo]�
► Press [Set] until the requested value is displayed (intervals from 1 hour
to 8 weeks)�
► Briefly press [Mode/Enter]�
► Press [Set] until [rES�T] is displayed�
► Briefly press [Mode/Enter]�
> The counter is reset automatically with the value now set�
10.4.6 Deactivate meter reset
► Select [rTo] and set [OFF]�
> The meter is only reset after overflow (= factory setting)�
Overflow: After the maximum value the meter is reset to 0�
10.4.7 Configure meter reset using an external signal
► Select [OU2] and [InD]�
► Select [Din2] and set the reset signal�
-[HIGH] = reset for high signal
-[LOW] = reset for low signal
-[+EDG] = reset for rising edge
-[-EDG] = reset for falling edge
LED 7 (→ 7 Operating and display elements) indicates the input
status also with an active external reset�
10.5 Settings for temperature monitoring
10.5.1 Configure limit value monitoring with OUT2
► Select [SEL2] and set [TEMP]�
► Select [OU2] and set the switching function�
-[Hno] = hysteresis function/normally open
-[Hnc] = hysteresis function/normally closed
-[Fno] = window function/normally open
-[Fnc] = window function/normally closed
► Select [SP2] and set the value at which the output switches�
► Select [rP2] and set the value at which the output resets�