IET Labs HARS-LX User And Service Manual Download Page 18

12

HARS-LX Series

12

Maintenance

5.4  Adjustments

If one or more resistances fall outside the limits, the 

associated resistor should be adjusted.  There is a 

trimming  network  provided  for  each  resistance  in 

these ranges.  These may be accessed by removing the 

housing and accessing the particular decade PC board.

Adjust the resistors from lowest to highest.  The order 

of decades does not matter.

Whereas it is possible to adjust any one resistance 

step, note that the 

n

th

 step of a decade is the sum of 

resistances 1 through 

n

, so that errors are cumula-

tive.  It is therefore recommended that whenever any 

resistance of a particular decade is adjusted, that all 

the resistances of that decade be tested and adjusted 

as required.

To adjust any of the resistances, the following should 

be observed:

• 

Adjustment considerations

• 

Adjustment procedure

5.4.1  Adjustment Considerations

Before adjusting or trimming any resistances, observe 

the following:

The equipment measuring the unit should conform to 

the guidelines provided in Section 5.3.

  

CAUTION 

The  HARS-LX  front  panel  GND 
binding post should be connected to 
the test instrument’s ground point.

5.4.2  Adjustment/Trimming 

Procedure

To trim or adjust any resistances, 

proceed as follows

:

1.  Stabilize HARS-LX at laboratory tempera-

ture of 23 °C for at least 8 hours.

2.  Remove the 4 housing screws from the sides 

of the instrument and remove the housing.

3.  Using  the  binding  posts  in  the  4-terminal 

connection, connect HARS-LX to a resis-

tance meter.

4.  Rotate all knobs approximately 10 times and 

be careful not to rotate the minimum setting 

decade beyond the stops.

Breaking in the switches helps elimi-

nate any residual contact corrosion 

resistance.

5.  Set all dials to 0, except for the minimum 

setting decade.

The minimum setting decade has a mini-

mum stop at position 1.

6.  Measure the minimum resistance setting

If an adjustment is needed, consult IET 

technical support.

7. 

Measure 1 mΩ, 10 mΩ, and 100 mΩ steps as 

applicable to the HARS-LX being trimmed. 

If adjustment is needed, proceed as follows:

• 

If the resistance is 

high

, reduce it by add-

ing a small bit of solder to an exposed 

portion of the wire

• 

If the resistance is 

low

, increase it by 

gently filing the wire in one spot.

8.  Measure the higher decades.  If an adjust-

ments is needed, proceed as follows:

Starting with the lowest setting, adjust the 

associated trim pots.

See Table 5-1 and Figure 5-1 for 

reference

9.  Replace the housing and reinstall the 4 hous-

ing screws.

  CAUTION 

Make these steps carefully and slowly, 
allowing the wire to cool down before 

measuring again.

Summary of Contents for HARS-LX

Page 1: ...ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HARS LX Laboratory Standard Decade Resistance Substituter User and Service Manual Copyright 2014 IET Labs Inc Visit www ietla...

Page 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Page 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IET shall...

Page 4: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Page 5: ...1 Thermal emf Considerations 5 4 4 Dial Setting 6 4 5 Power Considerations 6 4 6 Environmental Conditions 6 4 7 Switch Conditioning 6 4 8 Meter Shunt Applications 7 4 9 Kelvin Bridge Applications 7 C...

Page 6: ...tituter 1 Figure 2 1 Typical Operating Guide Affixed to Unit 3 Figure 4 1 Kelvin Bridge Connections 7 Table 5 1 Trimming Potentiometers 11 Figure 5 1 Typical Trimmer Board 11 Table 5 2 Replaceable Par...

Page 7: ...y feature solid silver alloy contacts and quadruple leaf silver alloy wipers which keep switch contact resis tance to under 1 m per decade and more impor tantly keep switch contact resistance reproduc...

Page 8: ...ermetically sealed low inductance Terminals Four 5 way gold plated tellutium copperbinding posts with low thermal emf and low resistance for four terminal Kelvin measurements plus one binding post con...

Page 9: ...00 1 211 0 M 4 100 100 HARS LX 4 1k 12 110 M 4 1 k 1 k HARS LX 4 10k 121 10 M 4 10 k none HARS LX 5 0 001 121 110 5 1 m HARS LX 5 0 01 1 111 10 k 5 10 m 10 m HARS LX 5 0 1 11 111 0 k 5 100 m 100 m HAR...

Page 10: ...100 short term average value Environmental Conditions Operating Temperature 0 C to 55 C Storage Temperature 40 C to 70 C Switch Type Multiple solid silver contacts dust tight diallyl phthalate body To...

Page 11: ...tandard temperatures near 23 C 3 3 Repackaging for Shipment If the instrument is to be returned to IET Labs contact the Service Department at the number or address shown on the front cover of this man...

Page 12: ...low resistance increments are important care must be taken in connecting to the terminals of the decade box In particular to keep contact resistance to a mini mum 4 terminal connections should be empl...

Page 13: ...0 Since the highest decade has an additional 11 po sition resistance values of over 1 2 times nominal maximum value can be obtained 4 5 Power Considerations To maintain the maximum possible accuracy a...

Page 14: ...resistance standard for all Kelvin bridge measurements When connected as shown in Figure 4 1 errors caused by lead and contact resis tances can be made negligible because they appear as part of the ge...

Page 15: ...igure 4 1 Figure 4 1 Rheostat Dial In order to eliminate contact resistance and thermal emf the HARS LX integrates the rheostat as shown in Figure 4 2 This way the wiper is in the low potential circui...

Page 16: ...from the side of the instrument and remove the housing 2 Remove any dust or debris using optical grade dry compressed air or a clean brush 3 Replace the housing and reattach the 4 hous ing screws The...

Page 17: ...he follow ing models HATS LR HATS Y SRL Series The 1 and 10 transfer standards are optional and are only used to take advan tage if desired of the adjustability of these two decades Precision resistan...

Page 18: ...trim or adjust any resistances proceed as follows 1 Stabilize HARS LX at laboratory tempera ture of 23 C for at least 8 hours 2 Remove the 4 housing screws from the sides of the instrument and remove...

Page 19: ...Designation All decades but 1 k Potentiometer Designation 1 k decade 1 T0 T1 2 T1 T2 3 T2 T3 4 T3 T4 5 T4 T5 6 T5 T6 7 T6 T7 8 T7 T8 9 T8 T9 10 T9 T10 11 T10 1 M steps only Table 5 1 Trimming Potenti...

Page 20: ...y Not Shown HARS 4000 LX 0 1 100 m step Decade Switch Assembly Not Shown HARS 4000 LX 1 1 step Decade Switch Assembly Not Shown HARS 4000 LX 10 10 step Decade Switch Assembly Not Shown HARS 4000 LX 10...

Page 21: ......

Page 22: ......

Reviews: