Maintenance
41
DE
EN
10
Maintenance
The monitor and the cuff are calibrated by the manufacturer for a period of two years. Maintenance (metrological
check) must be carried out in accordance with Directive 93/42/EEC every two years at the latest if the device is
used professionally. In certain countries, this requirement may be regulated by national laws or regulations.
The metrological check comes at an additional cost and can be carried out in Germany by either
IEM GmbH
, a
competent authority or by authorised maintenance services corresponding to the "Medizinprodukte-Betreiber-
Verordnung" (Medical Product Operator Regulations).
Other than the metrological check, no further electromagnetic compatibility-related maintenance is required.