Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
32
Recommended Separations from the
Aspect 100-4T G3
or the
Aspect ISR G3
are intended for use
in the electromagnetic environment in which radiated disturbances are controlled. The
customer or user of the
Aspect 100-4T G3
or the
Aspect ISR G3
can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the
Aspect 100-4T G3
or the
Aspect ISR G3
as recommended
below, according to the maximum output power of the Communications Equipment.
Table 4
Max Output
Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/3)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/3)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
D=(7/3)(Sqrt P)
0.01
.1166
.1166
.2333
0.1
.3689
.3689
.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333
17.
WARRANTY
Warranty details for IDI Products can be obtained directly from Image Diagnostics, Inc.
Image Diagnostics, Inc.
310 Authority Drive
Fitchburg, MA 01420 USA
Or call
IDI
at (978) 829-0009 or Toll Free (877) 304-5434.
The fax number for IDI is (978) 829-0027.
Recommended Separation Distances between portable and mobile RF
Communication Equipment and the Aspect 100-4T G3 or the Aspect ISR G3.
Equipment and Systems that are NOT Life-Supporting.