7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 39
LifeCare PCA
7.3.9.9
PATIENT
PENDANT
JACK
ASSEMBLY
REPLACEMENT
Recommended tools for this procedure are a 1/2 inch nutdriver, diagonal cutters,
and cable ties.
Replacement parts for this procedure are:
Assembly, Patient Pendant Jack
Tie, Cable
To replace the patient pendant jack assembly, see
,
and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosure assemblies as described in
3. Using the nutdriver, remove the retainer nut and flat washer that attach the patient
pendant jack to the rear enclosure.
4. Using the diagonal cutters, remove the cable ties that secure the patient pendant
jack assembly to the rear enclosure harness assembly.
5. Remove the spiral wrap and the patient pendant jack assembly.
6. Install the replacement patient pendant jack assembly in the exact reverse order
of removal. Secure the patient pendant jack harness to the rear enclosure harness
assembly with the spiral wrap and cable ties.
7. Reassemble the infuser in the exact reverse order of disassembly.
8. Connect the infuser to AC power.
To verify successful replacement of the patient pendant jack assembly, perform the PVT
in
CAUTION:
Do not bend the patient pendant cord near the patient pendant
or connector jack at an angle of 90
o
or greater.