background image

IDM

-60A     4-20mA

1. NOTICE

1. Please  check  the  packaging  and  appearance  after  you  received 
the product. 
2.Please  check  the  mode  and  specification  of  the  product.  Keep 
correct electrical connection and mounting method. 
3.Please installation WITHOUT power. 
4.During using,  please notice the technical and operation conditions: 
the temperature of medium, burst pressure, power supply. 
5.  This  product  is  a  precision  device,  DO  NOT  use  hard  objects  to 
tauch the diaphragm or disassemble. 
6.  During  installation,  do  not  install  or  disassemble  strongly, 

especially the installation of thread. 

7.  Please  install  or  disassemble  with  a  suitable  wrench,  Otherwise, 

the damage is not in the scope of warranty. 

8.  After starting the gauge,  it  is a normal phenomenon to take a few 

minutes to keep it stable and work normally. 

9. lf the product is out of order, please contact our after-sale technical 

staff. 
10.During  installation,  the  gauge  will  be  affected  by  installation 
stress. After installation, please  zero clearing. 

(!!)The  product  damage  caused  by  non-professional  operation 

according to the operation regulation is not the scope by the warranty. 

2. WARNING

1.When operation temperature is above 60 degree, please use forced 
fans  or cooling  machines,  but  do  not  let  the  cooling  air blow  to  the 
switch directly. 
2.The  installation,  commissioning  and  maintenance  of  this  product 
shall be conducted by qualified engineers or technicians member. 

3.lf the malfunction or exception of this product is likely to result in a 
major  system  accident,  please  sei  up  an  appropriate  protective 

circuit outside to prevent accidents. 
4.The company  does not undertake any  direct or indirect  lass  other 
than the product itself. 
5.The  company  reserves  the  right  to  change  the  manual  of  the 
product without notice. 

3. DESCRIPTION

This remote transmission pressure gauge is an economical digital 

pressure  transmitter  with  a  built-in  high-precision  pressure  sensor, 

which  can  accurately  display  the  pressure  in  real  time  and  transmit 

the  pressure  signal  remotely.  lt  has  the  characteristics  of  high 

accuracy and good long-term stability. 

This remote transmission pressure gauge adopts LCD display, with 

multiple  functions  such as  uni! switching,  zero  calibration,  full scale 

correction,  digital  filtering,  etc.,  simple  operation  and  convenient 

installation. 

This product can  measure water,  oil,  air and other  non-corrosive 

media to stainless steel,  and can replace the traditional pointer type 

remote pressure gauge or pressure transmitter. 

4. TECHNICAL SPECIFICATIONS

Pressure range:  -0.1-0.1 ... 0.6 ... 1 ... 1.6 ... 2.5 ... 6 ... 10 ... 60MPa 
Pressure type: gauge pressure, negative gauge pressure 
Overload pressure:  <10MPa 200%; :?10MPa 150% 

Accuracy grade:  0.5% FS 
Long-term stability:  0.5% FS/year 

Power supply: 12-28VDC (24VDC) 

Output signal: 4-20mA 

Load impedance:  S500O 

Operation temperature: -20-60"C 

Display screen: 4-digits LCD display 

Backlight color: white 

Measuring medium:  air,  water,  oil and other non-corrosive media to stainless 
steel. 

Case  material:  304 stainless steel 
Connection material:  304 SS (or customized) 

Product functions: uni! switching, clearing, digital filter adjustment 

5. KEY DESCRIPTION

008 

5.1  KEY DESCRIPTION 

Key 

Definition 

Description 

§

Set key 

Lang press to enter parameter setting state 

Short press to increase 

lncreasing key 

When changing the password,  lang press to 

<>hift 

© 

Decreasing key 

Short press to enter system function, 

adjust the number short press to decrease 

5.2  LCD DISPLAY 

Status 

Description 

Operation status  Display current pressure and uni! 

Short  press  the 

® 

key,  and  then  lang press 

for  5 

Zero correction  seconds  until the screen displays  C-L, and the reset  is 

completed (the reset must be operated without pressure) 

Note:  When the display pressure is  higher than  10% 

of the range, it cannot be cleared 

Setting status 

Refer to item 7 for general setting operation method 

Reviews: