2028P7 

Differential 
1bar 

 

Screw connector  
mbar / bar / PSI / kPa / inHg/ cmHg / inH20 / mH20 

-10°C to +50°C

Model:

 

Sensor type: 
Operating range: 
Connector Type:

  

Units: 

 

Operating Temperature:

  

Accuracy of Unit Only

:   

All ± 1 Digit ( 

0.1˚C Low scale; 1˚C High scale )

From -

10˚C to +50˚C: 0.15% 

of r 0.15% of full scale

Negative pressure: 0.3% of reading  + 0.3% of full scale

Particular care should be taken not to over pressure the device as this may rupture the sensor membrane. This is 
not covered by the manufacturers warranty. 

Note: The differential/gauge instrument will only measure relative pressures i.e positive pressure applied to the 
positive port or negative pressure applied to the negative port. 

Instrument Application 

The 2028P is suitable for the measurement of differential pressure. 

Instrument Operation 

The instrument requires two AA or equivalent cells. These are not supplied with the unit. 
When the batteries require changing the following symbol will be displayed:  
Please ensure the screws holding the battery compartment are firmly tightened when inserting or replacing batteries 
to ensure the IP67 rating is maintained. 
Press ON key to turn unit ON. Press OFF key to turn unit OFF. 
The unit will switch OFF automatically after 12 minutes, unless any key is activated. 
To disable automatic switch off, hold      Key when the unit is being switched on. This will disable the automatic 
switch-off function until the unit is switched off. 

Out of Range 

If pressure exceeds the scale range 

Out

 will appear on the display. 

Zeroing 

In order to achieve maximum accuracy, it is recommended to zero the instrument in the orientation it will be used 
before taking any measurements. 
Ensure both ports are open to atmospheric pressure, if indicated pressure is within 1.25% of full scale of the factory 
calibrated zero, press 

0.0

 key which will adjust the reading to zero. 

Hold Function 

To enable the 

Hold Function, press the “Hold”

 

 button, Press again to release the Hold Function. 

Hose fitting 

The input port(s) on the instrument are designed to accept a hose with 4mm Internal Diameter and 1mm wall. The 
hose is fitted by unscrewing the clamping nut, sliding the hose on to the stepped shaft and refitting the nut.  

WARNING 

 Maximum torque is 2Nm, only tighten the clamping nut and not fitted port otherwise this may shear 

the port. This damage will not be covered under warranty.  

Smoothing mode 

Press 

 key and the instrument continuously averages the last four measurements to give a more stable reading on 

the display. Press 

 again to return back to normal settings. 

Units of pressure 

Press 

 key to select required units of pressure. Chosen units will be saved when the instrument is switched off. 