![ICP DAS USA tM-TH8 User Manual Download Page 128](http://html1.mh-extra.com/html/icp-das-usa/tm-th8/tm-th8_user-manual_3864813128.webp)
tM-TH8 User Manual, Rev: A1.1 7MH-018-A11
128
3.7 Modbus Address Mappings
Address Description
Attribute
00257
Protocol, 0: DCON, 1: Modbus
R/W
00258
Modbus protocol, 0: RTU, 1: ASCII
00261
1: enable, 0: disable host watchdog
R/W
00267
Temperature scale,
0: Fahrenheit
, 1
:
Celsius
R/W
00269
Modbus data format, 0: hex, 1:
engineering
R/W
00270
Host watch dog timeout status, write 1
to clear host watch dog timeout status
R/W
00272
Write 1 to load factory calibration
parameters
W
00273
Reset status, 1: first read after
powered on, 0: not the first read after
powered on
R
10129 ~
10136
00129 ~
00136
Over/under range status of channel 0
to 7
R
30001 ~
30008
40001 ~
40008
Analog input value of channel 0 to 7 R
40257 ~
40264
Type code of channel 0 to 7
R/W
40385 ~
40392
Resistance offset of channel 0 to 7 in
0.1 ohms, valid range: 0 ~ 255
R/W
40449 ~
40456
Temperature offset of channel 0 to 7
in 0.1°C/°F, valid range: -128 ~ 127
R/W