PET-AR400 User Manual
V1.0.0
P.31
3.
Data Acquisition Modes
PET-AR400 offers two data acquisition modes
:
one is continuous real-time acquisition mode
triggered by software commands, and the other is N sample acquisition mode is activated when
the analog input exceeds the specified threshold value.
The table below shows the sampling frequency of each acquisition mode, including the total
number of channels sampled simultaneously and the maximum sampling rate for each channel.
A/D Trigger
Data Transmission
Total simultaneous
sampling channels
Maximum sampling rate
Software A/D Data
Acquisition Mode
Continuous
Transmission
1
12.8k/16k/32k/64k/128k Hz
2
12.8k/16k/32k/64k Hz
3/4
12.8k/16k/32k Hz
Analog Threshold
Trigger Mode
N Sample
Acquisition
1
12.8k/16k/32k/64k/128k Hz
2
12.8k/16k/32k/64k Hz
3/4
12.8k/16k/32k Hz
Tips & Warnings
The data in the table above were obtained through testing in a local area
network, and actual users need to consider factors such as network and
computer when using them.