System Setup
‘Speed’ can adjust the speed of the screen display data, the system is divided into four categories ‘low, standard, medium and high’. When
leaving the factory, the system default display rate is ‘standard’.
Maintenance
Password is needed to enter ‘Current calibration’, if there is no password, press ‘Enter’ directly.
To calibrate current:
eg: to calibrate current1 4mA;
After the user connects the multimeter to the internal ‘Cur 1+’ (output current one/positive) and ‘Com-’ (output current negative), if the
multimeter displays a mA value of 3.98mA, press direction key to modify until the current value displayed on the multimeter is the corre-
sponding milliamp value (4mA), the calibration is completed.
When testing the current1, the test leads of the multimeter should be connected to ‘Cur 1+’ & ‘Com-’, and when testing the current, it should
be connected to ‘Cur 2+’ & ‘Com-’, where ‘Cur 1+’ and ‘Cur 2’ ‘+’ is the positive pole, and ‘Com-’ is the negative pole, which is also the output
current common terminal.
After selecting the corresponding relay, press
[ENT] to hear a clicking sound, then the relay is judged to be normal.
Summary of Contents for ProCon D300 Series
Page 1: ...D300 Series pH ORP Digital Analyzer Controller Operating Manual ...
Page 7: ...D300 Series Operating Manual 7 Keypad Descriptions ...
Page 12: ...D300 Series Operating Manual 12 Measurement Settings Quick Start ...
Page 13: ...D300 Series Operating Manual 13 Calibration Settings Quick Start ...
Page 14: ...D300 Series Operating Manual 14 Alarm Settings Quick Start ...
Page 15: ...D300 Series Operating Manual 15 Current Settings Quick Start ...
Page 16: ...D300 Series Operating Manual 16 Communication Settings Quick Start ...
Page 17: ...D300 Series Operating Manual 17 System Settings Quick Start ...
Page 18: ...D300 Series Operating Manual 18 Maintenance Settings Quick Start ...