Medical

Medical

Edit Text

Outlines

Summary of Contents for HOTSPOT

Page 1: ...W Cardinal Dr Beaumont TX 77705 Phone 409 861 0788 Toll Free 866 861 0788 infraredcameras com sales infraredcameras com We Are IR PLEASE READ THIS MANUAL BEFORE SWITCHING THE UNIT ON IMPORTANT SAFETY INFORMATION INSIDE Medical Medical Edit Text Outlines ...

Page 2: ...CCURATE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITHOUT LIMITATION THOSE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT OR ARISING FROM A COURSE OF DEALING USAGE OR TRADE PRACTICE THE SOFTWARE LICENS...

Page 3: ...ator 7 Description of Symbols 8 Technical Specifications 9 Calibration Instructions 10 Intended Use 11 Measuring Operations 12 Advice 13 Maintenance and Cleaning 14 Troubleshooting 15 Conformity Standards 16 About ICI Contents 4 4 5 6 7 7 8 9 10 11 12 14 14 15 17 18 THESE SYSTEMS ARE INTENDED TO BE USED ADJUNCTIVELY NOT AS STANDALONE DEVICES ...

Page 4: ... thermometer to electric shocks Do not expose the thermometer to extreme temperature conditions of 50 C or 0 C Do not use the device in relative humidity higher than 85 Do not use the device near large electromagnetic fields such as found with cordless or cell phones Keep the device away from water and heat including direct sunlight Do not drop or throw the device Do not use the device if it is da...

Page 5: ...he surface temperature of an object This device is not intended as a substitute for a physician s consultation Should a problem occur with your device contact your retailer Do not attempt to repair the device yourself According to EMC standard the electronic products should be maintained Precise non contact measurements User selectable ºC or ºF Selectable Body and Surface temp Set Alarm value Memo...

Page 6: ...ormal Temperatures According to Measurement Method The temperature of the human body varies throughout the day It can also be influenced by numerous external factors age sex and thickness of skin 4 2 Normal Temperatures According to Age Measurement Method Normal Temp C Normal Temp F Rectal 36 6 to 38 97 8 to 100 4 Oral 35 5 to 37 5 95 9 to 99 5 Axillary 34 7 to 37 3 94 4 to 99 1 Ear 35 8 to 38 96 ...

Page 7: ...sor 2 LCD Display 3 Mode Selection 4 Down Button 5 Up Button 6 Mode Button 7 Measurement Trigger 8 Battery Cover 1 Surface mode Symbol 2 Body mode Symbol 3 Digital readout 4 Battery Symbol 5 The order number 6 Save data readout 7 Temperature Scale C Celsius F Fahrenheit 8 Buzzer symbol ...

Page 8: ...HOTSPOT IR NON CONTACT THERMOMETER USER MANUAL 8 7 Description of Symbols ...

Page 9: ...oss 400 g Net 172 g 8 1 Normal Conditions of Use 8 Technical Specifications 8 2 Measuring Range In Body Mode 32 0 C to 42 5 C 90 F to 108 F In Surface Temp Mod 0 C to 60 C 32 F to 140 F Accuracy 0 3 C 0 54 F Measuring Distance 5 cm 15 cm 2 5 9 Automatic Stop 7 seconds 32 C to 35 9 C 93 2 F to 96 6 F 0 3 C 0 5 F According to ASTM Standard E1965 1998 2003 36 C to 39 C 96 8 F to 102 2 F 0 2 C 0 4 F 3...

Page 10: ... the person If you get 37 C 98 6 F the HotSpot IR Non Contact Thermometer is properly set and ready for use If you get a lower or higher temperature 38 1 C 100 6 F for example you should re calibrate the HotSpot IR Non Contact Thermometer 9 2 Calibrate the Device To calibrate the device 1 Press the MODE button for 2 seconds The screen will display F1 2 Press MODE button again until you get F3 3 Pr...

Page 11: ...onds in order to first test the room temperature 4 Aim toward a person s inner canthus keeping 5 to 15 cm 2 to 5 9 distance between the thermometer and the person press the measuring key and the temperature will display immediately see the Figure 1 below for the positioning 5 When the room temperature is significant difference or there is perspiration on the forehead you can take the temperature b...

Page 12: ... Function To adjust the calibration of your thermometer 1 Press MODE button for 2 seconds screen will display F1 2 Press MODE twice to get F3 3 Press to increase the difference by 0 1 C 0 1 F 4 Press to reduce the difference by 0 1 C 0 1 F IN THE CASES OF SEASONAL OR ENVIRONMENTAL CHANGES A VERIFICATION AND ADJUSTMENT SHOULD BE CARRIED OUT THIS FUNCTION IS ONLY EFFECTIVE IN BODY MODE 11 4 Buzzer O...

Page 13: ...D Press or button to display the last temperature measurement When the device is powered off press and button together for two seconds to display the last temperature measurement Change the order number to 0 and press MODE button to delete all collection data 11 8 Changing The Batteries When the LCD screen displays the battery needs to be replaced To replace batter open the lid and change the batt...

Page 14: ...rgeable batteries Do not throw the device in a fire Do not expose the thermometer to extreme sunlight or water The infrared sensor is the fragile Be careful with it Clean the device with a cotton swab lightly moistened with 70 alcohol Do not clean the device with a corrosive detergent Keep the device away from water or other liquids Store the device in a dry environment Keep the device away from d...

Page 15: ...urface Temperatures If you are using Surface mode the temperature displayed is showing the ambient temperature and not the surface skin temperature measurement Message HI The analysis is above the measurement range selected either above 42 5 C 108 F in Body mode or below 60 C 140 F in Surface mode Message LO The temperature analyzed is under the measuring range selected either the measurement is l...

Page 16: ...struction prior to taking temperature measurements Temperature hampered by an air flux Make sure there is no air flux as this could interfere with the infrared system The measuring distance is too far Be mindful the measuring distance is 5 cm 15 cm 2 5 9 From high low temperature condition to room temperature Wait for 10 minutes after being in a room before taking body temperature measurements ...

Page 17: ...al standard Electromagnetic compatibility Requirements and test IEC 60601 1 2 2001 EMC Statement This device has been tested and homologated in accordance with EN60601 1 2 2007 for EMC This does not guarantee in any way that the device will not be affected by electromagnetic interference Avoid using the device in high electromagnetic environment The ELECTRICAL EQUIPMENT needs special precautions r...

Page 18: ... Institute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion proof systems Our knowledge and experience stems from years of using infrared imaging and temperature measurement instruments to provide solutions to managers engineers scientists inspectors and operator...