5.6.2
ProSense™
cryoablation system touch screen failure
If the touch screen shuts down, you will need to shut OFF the system with the
mechanical ON/OFF button and remove the plug from the power supply. This will end
the cryotherapy procedure.
Warning
In case of software crash, switch OFF the mechanical ON/OFF button and
unplug the electrical cable.
Call IceCure Medical for technical service before restarting the ProSense™
cryoablation system.
5.6.2.1
Emergency Stop button
The fastest way to end a procedure is to press the RED HANDLE button or EXTRACT on
the screen.
However, in case of electrical emergency, press the RED ROUND Emergency
Stop button located ON the right side of the main chassis. The Emergency Stop button is
designed for emergency shutdown of the unit. Pressing this button immediately turns
OFF the system.
Switch OFF the mechanical ON/OFF button and unplug the electrical cable.