Figure 99: Last Treatment screen after up to 3 Freeze steps
Replace (refill) the Dewar during a cryoablation procedure
Option to refill the Dewar is available at each
Thaw
step.
If the Freeze step is 5 minutes or longer, the system will request Dewar refill between
the steps.
It is recommended to have an additional full Dewar ready, in case you need it for a longer
freeze time.
•
At every
Thaw
, the system will automatically ask if you want to replace the Dewar
(see figure below):
Figure 100: Replace Dewar option question
•
Press
OK
to replace the Dewar.
Note:
If the Freeze step was shorter than 5 minutes,
Cancel
button will appear on the
screen.