![IceCure ProSense DMS-7063 User Manual Download Page 87](http://html1.mh-extra.com/html/icecure/prosense-dms-7063/prosense-dms-7063_user-manual_3378524087.webp)
IceCure™ Medical Ltd.
DMS-7063 rev. C
ProSense™
Confidential
User Manual
87
US User Manual
Cryoprobe
Diameter/ Gauge
Shaft
Tip ±3mm
Tip Shape
Ice Ball Shape
Cool Zone Center
From needle tip
[mm]
Part Number
3.4 mm / 10G
140
mm
Trocar
Elliptic
20 mm
FAP7200000
3.4 mm / 10G
185
mm
Pencil
Elliptic
20 mm
FAP7400000
3.4 mm / 10G
185
mm
Trocar
Elliptic
20 mm
FAP7410000
2.4 mm / 13G
124
mm
Trocar
Spheric
10 mm
FAP7600000
2.4 mm / 13G
134
mm
Trocar
Elliptic
14.5 mm
FAP7800000
•
Cryoprobe and matching introducers are marked in the same color.
Introducer
Shaft Length
Description
Tip Shape
Part Number
115 mm
Fit to needle diameter 13G (FAP7800000)
Trocar
FAC9000000
122 mm
Fit to needle diameter 10G (FAP7200000)
Trocar
FAC9100000
167 mm
Fit to needle diameter 10G
(FAP7400000 and FAP7410000)
Trocar
FAC9200000
*
IceCure™
Medical systems, needles, introducers, cryoprobes and accessories may not be available in all
regions.
Refer to Cryoprobe and Introducer IFUs for further details.
Figure 80: Introducer shaft length
The order in which you follow these steps is essential for maintaining sterility and
patient safety.
First,
insert the cryoprobe tip into the target tissue. Be aware of the markings on the
cryoprobe: the wide mark closest to the tip is the safety mark. In percutaneous
procedures
it must be completely
inside the tissue
to avoid skin burns. The rest of the
marks indicate depth of cryoprobe insertion: each mark equals one centimeter with
distinctive markings at 5 and 10 cm as shown in Fig. below.
Figure 81: Illustration of FAP7200000 cryoprobe Markings on the cryoprobe.