Caution
The system will not allow additional treatment when zero procedures are
left to maintenance. Make sure to call
IceCure™
Medical service
representatives in time.
Warning
The system and its accessories are not compatible with magnetic resonance
imaging.
Cyber security
The
ProSense™
Cryoablation System has no Bluetooth or WiFi devices installed and
cannot be connected to any networks or devices using Bluetooth or WiFi protocols.
Caution
Use USB2.0 flash drives only to export reports or log files. Importing any data
or software may corrupt the
ProSense™
Cryoablation System.
Do not connect any other USB devices to the
ProSense™
Cryoablation System
USB port.
Caution
Do not use USB extension cables to connect a USB flash drive to the
ProSense™
Cryoablation System USB port. Using USB extension cables may
cause electromagnetic emissions exceeding regulatory limits.
Caution
Do not connect Ethernet cables to the
ProSense™
Cryoablation System and
do not connect the system to any internal or external networks.
Connecting external equipment to the system can only be done by individuals
trained and certified by
IceCure™
Medical company.