Figure 101: Lowering the Dewar during Thaw screen.
The message "Please wait…Do not open the door yet!" will
appear.
•
The system will perform several steps, and then it will inform you that it is safe to
replace the Dewar and you will get the Replace Dewar screen (figure below).
•
Please insert a full Dewar in place of the old one and only then press
OK
to
continue.
•
At the same time, you will get an interactive update of the Thawing time.
Figure 102: Replace Dewar screen
•
After pressing the
OK
button, the system will proceed and prepare the new Dewar
that was inserted. The screen will show
"Proceeding, please wait… "
(figure below)