IceCure
™
Medical Ltd.
DSR3200000 rev. G
ProSense
™
Confidential
User Manual
102
Return to first page U.S User Manual
Figure 98: Last Treatment screen after up to 3 freeze cycles
5.3.8
Replace (refill) the Dewar during a cryoprocedure
Option to refill the Dewar is available at each THAW cycle.
Dewar refill should be done between any two freezes, if the first FREEZE
duration is greater or equal 5 minutes, and the total freeze duration
planned for both FREEZE procedures is greater or equal 10 minutes.
First, always have an additional full Dewar ready, in case you may need it for a
longer freeze time.
•
At every THAW, the system will automatically ask if you want to replace the Dewar
(see figure below):
Figure 99: Replace Dewar option question
•
If you press CANCEL, you will stay in the THAW screen.