![Icare HOME2 Instruction Manual Download Page 34](http://html2.mh-extra.com/html/icare/home2/home2_instruction-manual_3335059034.webp)
34
4. Pull the probe base off.
5. Insert a new probe base into the tonometer.
6. Put the collar back to the tonometer.
7. Turn the collar clockwise until it is firmly in place. Do not use excess
force.
Dispose of the used probe base. To order new probes or probe bases,
contact the organization from which you obtained the tonometer or Icare
USA.
14.2 Clean and disinfect the tonometer
WARNING!
Never immerse the tonometer in liquid. Do not spray, pour, or spill liquid on
the tonometer, its accessories, connectors, switches, or openings in the cover.
Immediately remove any liquid from the surface of the tonometer.
WARNING!
The probe base must be changed, not cleaned.
PRECAUTION!
Certain microbiological agents (for example, bacteria) can be transmitted
from the forehead or cheek support. To prevent this, clean the forehead and cheek
supports with disinfectant for each new patient.