33
14 Maintenance
WARNING!
The tonometer should only be opened by qualified iCare service personnel.
The tonometer does not contain any user-serviceable parts, apart from the batteries and
the probe base. The tonometer does not require any routine servicing or calibration
other than changing the batteries at least annually and the probe base every six months.
If there is a reason to believe that the servicing of the tonometer is necessary, contact
the manufacturer or local distributor.
WARNING!
The tonometer must not be repaired or re-assembled by any other than the
manufacturer or its authorized service center. If the tonometer is broken, do not use it.
Take it to an authorized iCare service center for repair.
WARNING!
To avoid possible damage, keep the tonometer out of the reach of children
and pets. The probe base, battery cover, screws, collar, and probes are small objects and
may be accidentally swallowed.
WARNING!
Do not change the batteries or the probe base when the USB cable is
connected.
WARNING!
Servicing or maintenance actions must not be performed while the
tonometer is in use.
14.1 Replace the probe base
WARNING!
The tonometer must be switched off when the probe base is changed.
WARNING!
The probe base must be changed, not cleaned.
PRECAUTION!
To ensure the correct functioning of the tonometer, change the probe
base every six months.
NOTE!
Change the probe base if the tonometer constantly prompts REPEAT or CHANGE and
changing the probe does not resolve the issue.
The probe base may function improperly if dirt or liquid gets inside it.
1. Switch off the tonometer.
2. Turn the probe base collar counterclockwise until it becomes loose.
3. Lift the collar off the tonometer.