Last update:
2022-05-31
www.ibspe.com
User Manual |
P6
9
Ungrounded targets
Ungrounded targets have the potential to inject noise into the sensor. If the output is unusually noisy, be sure the
target is grounded. On moving/rotating targets this can be accomplished with a small metal brush or thin piece of
metal which is connected to ground.
10
Multiple systems
All channels within the ECL150 are synchronized and can be used on the same target. When multiple ECL150s
are used with the same target, the systems must be synchronized for best performance. An interconnecting cable
connects between the Sync In and Sync Out connectors. Single systems do not come with these connectors
installed.
11
Setting bandwidth
The system is shipped with the bandwidth setting listed on the calibration certificate based on order
specifications. To change bandwidth, remove the I/O connector endplate and set switches according to the table
below. Bandwidth specifications are -10%/+30%.
SW1
SW2
Bandwidth
Up
Up
250 kHz
Down
Up
1 kHz
Up
Down
10 kHz
Down
Down
15 kHz